Trending Topic

23 mins

Trending Topic

Developed by Touch
Mark CompleteCompleted
BookmarkBookmarked
Luke G Qin, Michael T Pierce, Rachel C Robbins

The uvea is a vascular stratum that includes the iris, ciliary body and choroid. Uveitis is defined as inflammation of a part of the uvea or its entirety, but it is also used to describe inflammatory processes of any part of the eye, such as the vitreous or peripheral retina. The clinical taxonomy of uveitis […]

Solving the Pitfalls of Bimanual Phacoemulsification – Oval Instruments Do Save Energy

Fernando Araújo-Gomes
Share
Facebook
X (formerly Twitter)
LinkedIn
Via Email
Mark CompleteCompleted
BookmarkBookmarked
Copy LinkLink Copied
Download as PDF
Published Online: Jan 20th 2011 European Ophthalmic Review, 2007:39-41 DOI: http://doi.org/10.17925/EOR.2007.00.00.39
Select a Section…
1

Article

In physics, energy transfer per unit of time equals power, or the amount of work done. The use of fluid as a tool to carry out work is known as hydraulics. In hydraulics, the use of the kinetic energy of a moving fluid is known as hydrodynamics, and the use of the static pressure of a fluid contained in a closed system with a reservoir and piping is known as hydrostatics.

Hydrodynamics and Hydrostatics


In physics, energy transfer per unit of time equals power, or the amount of work done. The use of fluid as a tool to carry out work is known as hydraulics. In hydraulics, the use of the kinetic energy of a moving fluid is known as hydrodynamics, and the use of the static pressure of a fluid contained in a closed system with a reservoir and piping is known as hydrostatics.

Hydrodynamics and Hydrostatics
Phacoemulsification should be regarded as a complex system comprising a hydraulic machine with piping, providing work or power to a chamber – the eye. The work inside this chamber becomes a hydrodynamic process with fluid flow and pressure depending on the phaco machine itself. The whole system – made up of the machine and the eye – constitutes a hydrostatic power transmission system, with a reservoir for the fluid and tubing and a pump to drive the fluid. The difference lies in the pump – a vacuum pump, which is needed to provide the fluidics in the anterior chamber – and in the fact that the fluid is not re-used. Basically, however, the system uses the power transmission of the static pressure that stands inside the tubing.

From any handbook of physics we can learn that, since the connection between the elements is through the piping, the principal advantage of hydrostatic power transmission (see Figure 1) is that these elements can be positioned in whatever fashion is suitable for the machine manufacturer. Hydrostatic transmission also has the advantage of not depending on the size of components.

However, the main disadvantage of hydrostatic drive is that it may lose efficiency due to internal leakage, fluid-flow losses or mechanical frictionlosses (see Figure 2), all of which result in power being wasted. In addition, if for any reason any external leakage or fluid loss is apparent, more power is wasted and the system becomes more inefficient. Assuming that we are dealing with hydrostatic transmission, our aim in terms of phacoemulsification itself is to spare the energy, or power, lost through leakage. What happens, then, if our phacoemulsification is made bimanually with a sleeveless round needle?

Incisions – Size, Shape, Misconceptions and Frequent Mistakes
It is important, before going on to leakage, to understand common mistakes and misunderstandings surrounding bimanual phacoemulsification.

Size
Most irrigating choppers currently in use are 19G. Until now, irrigating choppers and phaco needles have been round. This has meant the introduction into the elastic tissue of the limbus of a circular instrument with a certain gauge through a linear incision. Since it is very common to read or hear of surgery performed using a 1mm incision, we will take this measurement as our example. We will consider a 19G instrument with a circular section of 1mm in outer diameter. It is this circle that needs to be introduced through a 1mm incision. Using the formula 2πr, for a diameter of 1mm we calculate a perimeter of 3.1416mm. As the limbus is vertically elastic, the incision will open equally up and down, making the size of the incision half of the perimeter just 1.57mm. This means that for a 19G instrument (1mm in diameter) the real size of the incision needed would be 1.57mm. The same considerations work for a 20G 0.9mm instrument, which would require an opening of 1.41mm. This is a lot greater than 1mm; therefore, it is probable that when we hear or read about 1mm incisions, the author is actually referring to 1mm instruments. These disparities are due to the fact that the diameter of the instruments is reported to be equal to the size of the incisions. This in itself is based on the assumption that the incision fits perfectly around the circular instruments. This is the second mistake – as well as the size of the incision being misjudged, the shape of the incision is unclear.

Shape
As previously noted, conventional phaco instruments are round, meaning their section is a circle. In a circle the diameter is constant at any point. However, we must understand that in phaco surgery we are dealing with the perimeter of the incision, not with the diameter of the instrument. A knife produces a linear opening, and the elasticity in the limbus or cornea does not allow us to transform this line into a circle, even if the incision is forced. Forcing it would make the surgical incision into a traumatic wound, which would be out of the question. The lack of elasticity in the vertical plane of the limbus means that the incision must be widened horizontally to allow the smooth entrance of the round instrument without lacerating. The best achievable result, therefore, using a gentle and precise technique, is an oval opening. The incision, therefore, does not match the circular shape of the instrument. When we consider the size of the incision we must envisage an oval and forget the circle, which is the source of all misunderstandings. Furthermore, instead of one single diameter, we now have one minor vertical axis and one major horizontal axis. In other words, if vertical elasticity allows the entrance of the diameter of a round instrument, this will be the minor axis of the oval opening; horizontally, due to the lack of elasticity, we must widen the incision until a forceless entrance is possible – this will be the larger or major axis of the oval opening, and approximately the real size of the incision. When a round instrument is used, since the diameter is constant, it is in this horizontal plane that a space for leakage is created. However, as with the diameter of the circle, what we must consider for sizing the incision is not the size of the major axis or the minor axis, but half the perimeter of the oval (see Figures 3–6).

The True Shape and the Real Incision Size
It is concluded, therefore, that with round instruments (choppers and phaco needles) a circle is introduced into an oval opening. With conventional coaxial phaco, however, the needle is surrounded by a silicone sleeve, and this is the real difference between the techniques. This silicone sleeve matches the shape of the incision, while the bare instrument does not, as explained above. With the sleeve, therefore, excessive leakage does not occur. Furthermore, the second hand holds a thin conventional chopper through an opening smaller than 1mm, with leakage leaving the system through the ‘escape valve’ that every closed system needs. However, removing the sleeve from the phaco needle and using a round irrigating chopper in the second hand reveals space around the instrument, allowing excess leakage. The second hand, which holds the irrigating chopper, experiences the same problem. This creates a level of leakage that is dangerous and compromises the system. Bearing in mind the flow losses in the hydrostatic system, this is the main reason why chambers collapse.

This raises doubts about the benefits of using the bimanual technique. Bimanual, however, has several advantages over coaxial. First, the bimanual technique separates irrigation from aspiration, which is why it is referred to as biaxial surgery. The separate irrigating axis, in the second hand, provides the surgeon with another tool. By freely directing the fluid in several directions during the procedure, the surgeon can move the nuclear fragments from the angle to the aspirating tip, use the flow as a holder for the posterior capsule and even stabilise a floppy iris. However, directing the fluid is possible only if the irrigating chopper has a frontal, open-ended irrigation. Bimanual also enables the use of a truly ultra-small incision. Despite new evolutions in conventional phaco tips, I believe that only a sleeveless needle provides a truly small incision – 1.4mm or less. Finally, bimanual offers advantages in terms of fluidics. As we are working with ultra-small incisions, if we solve the leakage problem, we will achieve a much more efficient relationship of inflow to outflow in the anterior chamber, which will result in a safer and more predictable surgery. This is the key point on which everything else depends. It is because of these advantages that we must go on trying to solve the leakage problem.

The True Shape – Saving Energy by Reducing Leakage
Surgeons trying to compensate for this excessive leakage tend to raise the fluid bottle levels to around 1m, slow the pump velocity, lower the aspirating settings or use higher sizes for the chopper in order to achieve more flow. External pumping systems are also occasionally used in order to make the inflow more efficient. In spite of all these adaptations, the anterior chamber is barely stabilised and is placed under too much stress. So far, no advantages are evident over the coaxial technique, which is frustrating. To summarise, we hope to preserve the advantages of an ultra-small incision and the benefits of separate irrigation and aspiration. However, much leakage is evoked, causing the chamber to collapse. The problem of excessive leakage depends on the shape of the instruments, and the gain in efficiency on the size of the incision. The only way to solve the shape conflict, if the incision is definitely oval, is to use oval instruments.

Take the example of a 20G implement (see Figure 6). The area of a 20G circle is 0.5. To achieve a 0.5 oval, given 1mm as the major axis, and using the above formula, the minor axis must be 0.6mm. Taking 1mm and 0.6mm as the axes, the perimeter will be 2.65mm. The incision should therefore be half the perimeter – 1.32mm. Note that, even with a 1mm major axis, the incision will never be 1mm. Ultra-small-incision cataract surgery (USICS) is my personal designation for an effective bimanual cataract surgery using an incision no greater than 1.5mm. Incisions under 1.5mm with minor leakage, and a well-fitting instrumentwith an oval opening, produce better fluidics inside the anterior chamber, hence a more stable and effective surgery. Using the half-perimeter of the instrument as the real size of the incision, and using oval shapes to provide a better fitting, are the key issues to resolve the pitfalls of bimanual sleeveless phaco surgery.

The A Gomes Ultra-small-incision Cataract Surgery Kit
Based on these conclusions, I designed an oval irrigating chopper and, most importantly, an oval needle; without this oval needle, the whole technique would be ineffective. I selected 1.2mm and 0.6mm as the axes for the oval instruments, with 20G able to pass through 1.4mm, in order to retain the concept of a surgery with an incision smaller than 1.5mm. Buerki Innomed, in Switzerland, manufactured these instruments and they have since been put to use on a regular basis. As well as the chopper and needle, there is one irrigation handpiece and one aspiration handpiece, which are both oval. The set was named the A Gomes USICS Kit. After more than 750 consecutive cataracts, ranging from + to ++++, the oval instruments have proved to work perfectly. They have been used with Pentasys, Millennium and Sovereign machines, without any of the regular settings needing to be changed, and without the occurence of any flat chambers. They work with pump velocities of up to 40cc/mn, without the bottle increasing to more than 75cm (25 inches) high.

Since I consider bimanual surgery to be a technique involving incisions of less than 1.5mm, and as we now have some intraocular lenses (IOLs) that go through incisions of 1.75mm, I assume that with these instruments we can perform the whole surgery, from beginning to end, with an incision of no greater than 1.8mm. This surpasses the currently accepted standards, which define bimanual surgery as surgery with incisions under 2mm. For softer cataracts or for almost clear lenses, as for presbyopic lens exchange, I am now using the same concept but with 21G instruments, which work through 1.2mm incisions.

Conclusion
Bimanual cataract surgery has many advantages over conventional phaco surgery. It provides the surgeon with an independent irrigation line, allowing biaxial surgery. It allows the benefits of a truly small incision. Once oval instruments are put to use it will mean a better relationship of with a smaller incision the sleeve must be removed, and in doing this too inflow to outflow. Altogether, these elements form a more closed system, which enables a very safe and predictable surgery. The use of this small-incision technique means the surgeon can insert the latest developments in small-incision IOLs.

2

Article Information

Received

2011-01-20T00:00:00

3

Further Resources

Share
Facebook
X (formerly Twitter)
LinkedIn
Via Email
Mark CompleteCompleted
BookmarkBookmarked
Copy LinkLink Copied
Download as PDF

This Functionality is for
Members Only

Explore the latest in medical education and stay current in your field. Create a free account to track your learning.

Close Popup